| Performance | |||||
| Measuring range | 0 – 10 mill cells/ml | ||||
| Performance range | 0.1 – 1.5 mill | ||||
| Repeatability* | CV < 6% 100-299k SCC/ml CV < 4% 300-499k SCC/ml CV < 3% 500-1500k SCC/ml |
||||
| Accuracy |
< 10% relative mean diff. from DMSCC (Direct Microscopic Somatic Cell Count) |
||||
| Carry-over | < 1% relative | ||||
| Sample types | Raw milk (cow, goat, sheep, buffalo) |
||||
| *CV = Coefficient of variation (STDev/AVG) x 100. (STDev = Standard deviation. AVG = Average) |
|||||
| Application data | |||||
| Analysis capacity |
100, 200, 300, 400, 500 or 600 samples per hour | ||||
| Sample intake | 2.5 ml (programmable 2.0 – 5.0 ml) |
||||
| Required sample temperature | 37- 42 °C (86-107.6 F) | ||||
| Working factor | 100 | ||||
| Standards and approvals Fossomatic™ 7 is CE-labelled and complies with the following directives and regulations: • EMC (ElectroMagnetic Compatibility) Directive 2014/30/EU • LVD (Low Voltage) Directive 2014/35/EU • Machinery Safety Directive 2006/42/EC • Regulation (EC) 1272/2008 on classification, labelling and packaging of substances and mixture, CLP (EC) • WEEE Directive 2012/19/EU • Packaging and packaging waste Directive 94/62/EC • REACH 1907/2006/EC |
|||||
|
Fossomatic technology complies with: |
|||||
| Specifications MilkoScan™ 7 RM Most of the calibrations are using multiple wavelengths selected freely from the entire Mid-IR spectrum in order to optimise robustness and accuracy. Compared to traditional filter calibrations, they are called full spectrum calibrations. |
|||||
| Performance Carry-over for all components <1% relative |
|||||
| Component |
Measuring range | Performance range | Repeatability | Accuracy Bulk | Accuracy Single cow |
| Fat |
0-15% | 2-15% | CV < 0.5% | CV < 1.0% | CV < 1.5% |
| Protein | 0-10% | 2-10% |
CV < 0.5% | CV < 0.9% | CV < 1.5% |
| Lactose | 0-10% | 2-10% | CV < 0.5% | CV < 0.9% |
CV < 1.5% |
| Solids | 0-20% | 2-20% | CV < 0.5% | CV < 1.0% | CV < 1.5% |
| Urea patented | 10-100 mg/dl | 10-100 mg/dl | Sd < 1.5 mg/dl | Sd < 3 mg/dl | Sd < 3.5 mg/dl |
| Citric acid | 0.1-0.5% | 0.1-0.5% | Sd < 0.005% | Sd < 0.01% | Sd < 0.015% |
| FPD (screening) | 400-600 m°C | 450-550 m°C | Sd < 0.5 m°C | Sd < 4 m°C | N/A |
| Novel parameters | |||||
| Fatty acids profile |
|||||
| Ketosis screening (BHB, acetone) | |||||
| Untargeted screening raw milk (adulteration) | |||||
| Application data |
|||||
| Analysis capacity |
100, 200, 300, 400, 500 or 600 samples per hour | ||||
| Sample intake | 5 ml | ||||
| Required sample temperature | 37 - 42°C | ||||
| Performance specifications | Full spectrum calibrations | ||||
| Optimize your instrument performance |
|||||
| Digital services |
Streamline your operations and improve your profitability with our FOSS IQX™ digital solutions |
||||
|
Standards and Approvals MilkoScan™ 7 RM is CE-labelled and complies with the following directives and regulations: • EMC (ElectroMagnetic Compatibility) Directive 2014/30/EU • LVD (Low Voltage) Directive 2014/35/EU • Machinery Safety Directive 2006/42/EC • Regulation (EC) 1272/2008 on classification, labelling and packaging of substances and mixture, CLP (EC) • WEEE Directive 2002/96/EC • Packaging and packaging waste Directive 94/62/EC • REACH 1907/2006/EC The MilkoScan™ 7 RM techniques comply with: • ISO 9622 / IDF 141:2013 • AOAC official method 972.16 By using wavelengths from the entire Mid- IR spectrum for each component, calibrations are optimised in terms of robustness and/or accuracy (temperature, homogenization and humidity) |
|||||