System Description
WineScan™ SO2 consists of the analyser and Foss Integrator software. Options for WineScan™ SO2 include the possibility to upgrade with colour (VIS) module and to automatic version with XY Autosampler. |
Flex*
Finished wine: (Ethanol, glucose/fructose, malic acid, volatile acid, total acid, pH, free sulphur dioxide, total sulphur dioxide)
and
Must: (pH, malic acid, tartaric acid, total acid, brix, density, free sulphur dioxide, total sulphur dioxide)
Auto*
Finished wine: (Ethanol, glucose/fructose, malic acid, volatile acid, total acid, pH, free sulphur dioxide, total sulphur dioxide)
More calibrations are available. A large number of parameters can be analysed simultaneously, and the number of measure profiles that can be set-up is unlimited.
|
Without SO2 analysis switched on:
Analysis time
|
30 seconds (excl. SO2, see below) |
Carry-over |
<1% |
Sample Temperature |
5 - 35ºC |
Sample Volume |
Programmable 4 - 25 ml, standard volume is 7 ml for Flex and 8 ml for Auto version |
Optical System
|
Hermetically sealed, humidity control |
Cleaning |
Automatic and programmable |
Calibration routines |
Slope & intercept adjustment |
Options in WinISI™ SW package |
PLS (Partial Least Squares) and modified PLS calibrations and PCA (Principal Component Analysis). Flexible selection of spectral intervals. |
With SO2 analysis switched on
Analysis time
|
150 seconds (options for fast hydrolysis time of 86 seconds)
|
Carry-over
|
< 2% |
Sample Temperature |
5 - 35ºC |
Sample Volume |
Fixed 4.2 ml to be added (total 8.2 - 29.2 ml) |
Hydrolysis agent |
4 ml 25% Phosphoric acid |
Optical System |
Independent detector, hermetically sealed, humidity control |
Cleaning |
Automatic and programmable |
Calibration routines
|
Slope & intercept adjustment |
WinISI™ SW options |
no SO2 - 32 liquid parameters available |
FTIR Calibrator |
FTIR Calibrator software for PLS calibration development |
* WineScan Auto is configured with an autosampler for the busy laboratory, WineScan Flex has manual sample intake.
|
|
|
Installation requirements
|
Power supply |
100 - 240 VAC ±10% – 50 - 60 Hz |
Power consumption |
Max. 600 VA during measurement, 200 VA in standby |
Ambient temperature |
5 - 35ºC |
Ambient humidity |
< 80% RH, cyclic up to 80% RH when going from low to high ambient temperature |
Weight |
89 kg for WSC Flex; 97,4kg incl. XY Autosampler |
Dimensions (H×W×D) |
54×88×47.3 cm (excl. PC) |
Environment |
For best performance, place the instrument on a stable surface away from excessive and continuous vibration |
Degree of protection |
IP43 (IP43 PC is optional)
|
Noise Level |
< 70 dB |
Fuse |
T 10.0 A |
Installation category |
II |
Pollution degree
|
2 |
Altitude |
< 2000 m |
XY Auto Sampler |
|
Power supply |
100 - 240 VAC ±10% – 50 - 60 Hz
|
Weight |
8.4 kg |
Dimensions (H×W×D) |
61 × 33 × 50.8 cm (with sample probe) |