Analysis frequency | Real time: Average analysis time per result 2 - 3 seconds | ||
Wavelength range | 1100 - 1650 nm | ||
Detector | InGaAs Diode Array | ||
Spectral dispersion InGaAs Diode Array detector | 1,1 nm/pixel | ||
Process line interface | Sapphire; Diameter 45 mm, thickness 12 mm, with food grade FFPM O-ring seal | ||
Product temperature |
Max 150°C (302°F) | ||
Product pressure | Production pressure < 21 bar (< 305 PSI). Shock pressure < 50 bar (< 725 PSI) |
||
General | |||
Technology | NIR technology | ||
Software package | ISIscan™ NOVA for instrument control | ||
Wavelength accuracy | <0.5 nm | ||
Wavelength precision | < 0.02 nm | ||
Wavelength temperature stability | < 0.01 nm/°C | ||
Spectral Noise | < 60 micro AU | ||
Vibrations - require optical fiber fixation |
0.4 Grms |
||
Ambient operating temperature | ProFoss™ 2 -5°C to 40°C (23 °F to 104 °F), cooling with a compressed air line allows use up to 65°C (149 °F) |
| | |
ProFoss™ 2 Ex ATEX / IECEX: 0°C to 50 °C (32 °F to 122 °F). cULus: 0°C to 40 °C (32 °F to 104 °F) |
Pressurised air – cooling (Amb. Temp. 45 - 65°C) | Cooling air Flow rate minimum 5 l/min, >99.9 % water free, >99.9 % free of oil and fine particles down to 0.3 μm | ||
Ambient humidity |
< 90% RH | ||
Dimensions (W x D x H) | w x h x d = 420 x 420 x 135 mm ( 16.5 x 16.5 x 5.3 inches) + brackets to hold the unit | ||
Weight | 25 kg (55.1 lb) | ||
Cabinet / Housing materials | 1.5 mm (lid 2.5mm) Stainless Steel EN 1.4301 (SS2333) | ||
Mechanical environment |
Process control equipment | ||
Degree of protection | ProFoss™ 2: IP69* | | | ProFoss™ 2 Ex: IP6X |
Approvals |
ProFoss™ 2: CE | | | ProFoss™ 2 Ex: CE, ATEX, IECEx and cULus certified (Dust explosion approved) |
Hygiene | 3-A hygiene and EHEDG hygiene certified | ||
Communication | KEPServerEX (Ethernet, Analogue Profibus/Profinet) to PLC/SCADA; FOSS IQX™ | ||
Network | High quality, shielded LAN cable; minimum category 5e. RJ 45 (IP 67) LAN connections |
||
Power supply | 1 phase, 100-240 VAC (max ±10 % of the rated voltage), max. 40 VA, 50 - 60 Hx | ||
Operation |
Indoor use or outdoor shielded from rain and direct sunlight |
* IP69 is the highest protection for dust entering the unit. IP69 means protected against the effect of high-pressure water and/or steam cleaning high temperature.