| Products measured and product presentation | ||
| Feature | Specification | |
| Dimensions (WxDxH) | 375 x 490 x 300 mm | |
| Weight | 27 kg | |
| Degree of protection | IP 65 | |
| Measurement mode | Reflectance or transflectance (for liquids) | |
| Wavelength range | 400 – 2500 nm | |
| Detector | Silicon (400 – 1100 nm), Lead Sulphide, PbS (1100 – 2500 nm) | |
| Optical bandwidth | 8.75 +- 0.1 nm | |
| Data resolution | 0.5 nm | |
| Number of data points | 4200 | |
| Number of subsamples | Default: 7 for small cup; 8 for large cup | |
| Self test | 12 minutes (variable) | |
| Analysis time | <1 minute for 32 scans (8 subsamples, 4 scans per subsample) | |
| Installation requirements | ||
| Voltage supply | 100-240 V AC*, frequency 50-60 Hz, Class 1, protective earth | |
|
Ambient temperature |
5 - 40oC | |
| Storage temperature | -20 to 70oC | |
| Ambient humidity | <93% RH | |
| Mechanical enviroment | Stationary during use | |
| EMS enviroment | Laboratory use, industry requirements | |
| *Mains supply voltage fluctuations not exceeding ±10% of the rated voltage. | ||