Digestor 2508 | Digestor 2520 | DT 208 | DT 220 | |
Tubepositions | 8 tubes | 20 tubes | 8 tubes | 20 tubes |
Automation | Automated | Automated |
*Manual |
*Manual |
Tube volume | **100 ml, 250 ml and 400 ml | 250 ml and 400 ml |
**100 ml, 250 ml and 400 ml | 250 ml and 400 ml |
Options | Rack system | Rack system | Rack system | Rack system |
Auto lift | Auto lift | |||
Exhaust | Exhaust |
Exhaust |
Exhaust |
|
Scrubber | Scrubber |
Scrubber |
Scrubber |
|
* time and temperature |
Temperature range |
100°C – 440 °C | |
Temperature setting repeatability | 1°C |
|
Temperature stability at 100°C | ± 5°C | |
Temperature stability at 400°C | ± 2°C | |
Heating up time 20 to 400 °Cat 230 V | DT 208: ~35 min DT 220: ~40 min |
|
Installation requirements | DT 208 Digestor | DT 220 Digestor |
Power supply | 100 – 115 V, 50 – 60 Hz 200 – 240 V, 50 – 60 Hz |
200 - 240V, 50 – 60 Hz |
Power consumption | 1100 W | 2300 W |
Weight net |
10 kg |
18 kg |
Dimensions (W x D x H) | 300 × 440 × 140 mm |
300 × 600 × 140 mm |
Water supply |
3 -5 l/min for 5 minutes, then 1 l/min* |
10 - 12 l/min for 5 minutes, then 3-5 l/min* |
Ventilation requirement | Exhaust system and fume hood | Exhaust system and fume hood |
*Not needed if scrubber is used |