| Feature | Specification | |
| Dimensions (w x d x h) | 375 x 490 x 300 mm |
|
| Weight |
27 kg | |
| Degree of protection | IP 65 | |
| Measurement mode | Reflectance or transflectance (for liquids) | |
| Wavelength range |
400 - 2500 nm | |
| Detector |
Silicon (400 - 1100 nm), Lead Sulphide, PbS (1100 - 2500 nm) | |
| Optical bandwidth | 8.75 ± 0.1 nm | |
| Data resolution | 0.5 nm | |
| Number of data points |
4200 |
|
| Number of subsamples | Default: 7 for small cup; 8 for large cup | |
| Self test | 12 minutes (variable) | |
| Analysis time |
<1 minute for 32 scans (8 subsamples, 4 scans per subsample) |
|
| Installation requirements | ||
| Voltage supply |
100-240 V AC*, frequency 50-60 Hz, Class 1, protective earth |
|
| Ambient temperature | 5 - 40°C |
|
| Storage temperature |
-20 to 70°C |
|
| Ambient humidity |
< 93% RH |
|
| Mechanical environment |
Stationary during use |
|
| EMC environment |
Laboratory use, industry requirements |
|
| *Mains supply voltage fluctuations not exceeding ±10% of the rated voltage. |
||
| Optimize your instrument performance |
||
| Digital services |
Streamline your operations and improve your profitability with our FOSS IQX™ digital solutions |
|
| Service agreement |
With a FOSS service agreement you will secure operational uptime, regulatory compliance, and analytical accuracy |
|